500 PRO USER MANUAL 25
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Neuroprobe
®
500 Pro Program Menu
Infrared
SLD Infrared Pads are active when the treatment timer is on. Output is cycled at 45.4 KHz, 62.5 KHz and 83.3 KHz
with 1 min per frequency, and at a 50% duty factor.
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Polarity:
+/-
+/- or + or -
Treatment time:
15 minutes
0
– 99 Minutes
Sensory - Motor
The Sensory - Motor modality contains five pre-programmed IFC modes and two pre-programmed Probe modes. In
IFC mode, output is provided through the output A-B jacks. The phase of the channels is indicated by the color of
the tip pins. The black wires are of the same phase and the red wires in the opposite phase relative to each channel.
All variable parameters may be altered during operation.
I. SENSORY - MOTOR PROBE TREATMENT PROTOCOLS
The Sensory – Motor protocols contain two pre-programmed probe protocols. All parameters may be varied during
operation. The following probe programs are available:
PROBE SENSORY:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
Burst frequency:
100Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0
– 50%
Burst scan time:
10 Seconds
0
– 20 Seconds
Modulation:
ON
ON or OFF
Stim time:
60 Seconds
0
– 999 Seconds
Waveform:
Burst Alternating Current
Output:
0 mA
0
– 1.8 mA
PROBE MOTOR:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
Burst frequency:
8.5Hz
.0 Hz to 999 Hz
Burst rate scan:
50%
0
– 50%
Burst scan time:
10 Seconds
0
– 20 Seconds
Modulation:
ON
ON or OFF
Stim time:
60 Seconds
0
– 999 Seconds