NEUROPROBE
®
500 PRO USER MANUAL
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
PADS SEQ MOTOR SENSORY:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
Burst frequency:
2 - 15 - 100 Hz
Burst rate scan:
20%
0
– 50%
Burst scan time:
10 seconds
0
– 20 Seconds
Beat:
Sweep:
Rate Scan Time:
Modulation:
ON
ON or OFF
Vector:
Auto Intensity:
0%
0
– 20%
Treatment time:
30 minutes
0
– 99 Minutes
Waveform:
Burst Alternating Current
Output:
0 mA
0
– 99 mA
Probe Noxious
The Noxious Probe mode contains one pre-programmed mode. All parameters may be varied during operation and
saved by the user. The following program is available in Noxious Probe mode:
PROBE NOXIOUS
Sensory / Noxious
Waveform
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Pulse rate:
4.0 Hz
0.1
– 15.0 Hz
Polarity:
+/-
+/- or + or -
Stim time:
60 Seconds
0
– 999 Seconds
Waveform:
Monophasic or Biphasic square wave
pulsed current
Output:
0 mA
0
– 1.5 mA
Nerve Block
The Nerve Block modality contains one pre-programmed Pad mode and one pre-programmed Probe mode. In Pad
mode, output is provided through the output A-B jacks. The phase of the channels is indicated by the color of the tip
pins. The black wires are of the same phase and the red wires in the opposite phase relative to each channel. All
variable parameters may be altered during operation.