15
Impella 5.5
®
with SmartAssist
®
Circulatory Support System
SYMBOLS
Caution; follow instructions for use
Defibrillator-proof type CF equipment
Keep dry
+10°C
+30°C
Storage temperature
(e.g. 10°C to 30°C)
Declares conformity with Directive
93/42/EEC for medical devices or with
Regulation (EU) 2017/745 on medical
devices, and with Directive 2011/65/EU
on the restriction of the use of certain
hazardous substances in electrical and
electronic equipment
2019-11-06
Date of manufacture (eg, 2019-11-06)
Protect from sunlight
Symbol for lot designation; the
manufacturer’s lot designation must be
stated after the LOT symbol
123456
Abiomed part number
(eg, part number 123456)
123456
Manufacturer’s serial number
(eg, serial number 123456)
2019-11-06
Use-by date
(eg, use before 2019-11-06)
Do not reuse
Sterilized using ethylene oxide
Medical Device
Do Not Flush
Do NOT flush
Glucose
Use glucose in the purge fluid
IMPELLA 5.5 CATHETER PARAMETERS
Speed range
0 to 33,000 rpm
Power consumption
Less than 1.07 W
Voltage
Max. 18 V DC
Flow-Maximum
5.5 L/min
Purging the Impella 5.5 Catheter
Recommended purge fluid
Glucose concentration
Purge pressure
Purge flow
5% glucose solution in water
with heparin concentration of
25 or 50 IU per mL
5% to 20%
300 to 1100 mmHg
2 to 28 mL/hr
Maximum duration of use
EU
Reliability
Up to 30 days
91.4% with 80% confidence at
30 days based on chi-squared
test
Dimensions of Impella 5.5 Catheter
Length of invasive portion (without catheter)
Diameter (Motor)
Diameter (Cannula)
114 mm
18 Fr (6.3mm)
21 Fr (7 mm)
Classification per DIN EN 60601-1
Protection class I, degree of
protection:
CF (Automated Impella
Controller and Impella 5.5
Catheter)
Classification per directive 93/42/EEC
Class III
Latex free
Yes
E
NG
LI
SH