5
Impella 5.5
®
with SmartAssist
®
Circulatory Support System
6.
Connect and tighten the yellow luer on the purge tubing to the Impella
Catheter sidearm if not already complete.
7.
When the controller detects that the luer is connected, it
automatically begins priming the purge lumen.
8.
Snap the purge clip (located on the pressure reservoir of the clear
sidearm) to the connector cable
9.
Once the purge lumen is primed, the controller automatically advances to
the next screen.
10.
The first step on the next screen prompts you to enter the purge
fluid information.
ENTER PURGE FLUID DATA
1.
Enter the purge fluid information.
2.
To select the default values displayed on the screen, press the
ACCEPT
soft button. This will select those values and automatically
advance to the next screen.
Note:
The default purge fluid values will
be the purge fluid values from the last Case Start performed on a
given Automated Impella Controller.
3.
To change the purge fluid information, press the
EDIT
soft buttons,
scroll to the appropriate item and push the selector knob to select it
or use the white arrow soft buttons. Then scroll through the values
and push the selector knob or press
SELECT
to make a new selection.
Press the
DONE
button to finish editing. The controller will use the
default values if no other selections are made.
• Purge fluid can be set to 50 mL, 100 mL, 250 mL, 500 mL, or
1000 mL.
• Glucose (Dextrose) concentration can be set to 5%, 10% or 20%.
• Heparin concentration can be set to 0 IU/ml, 5 IU/ml, 6.25 IU/ml,
10 IU/ml, 12.5 IU/ml, 20 IU/ml, 25 IU/ml, 40 IU/ml, 50 IU/ml.
SECURE THE PURGE TUBING
1.
To complete the setup, connect the purge tubing to the white
connector cable by pushing the purge tubing into the clips attached to
the white connector cable.