Type overview, technical data
54
Sensyflow iG, Version 1.00
42/14-39 EN
Type
Sensyflow iG
Sensyflow iG-Ex
Operating pressure
4
×
10
6
Pa (40 bar)
Operating temperature of
medium
(Transducer)
Standard range: -25...+150 °C
Extended range: -25...+300 °C
In preparation
Ambient temperature:
Evaluation electronics
Without display
With display
-25...+65 °C
-25...+50 °C
Other ambient temperatures on request
Storage temperature
-25...+85 °C
Degree of protection
IP 67 (IP 65 for remote version transducer)
Recommended
installation requirements
According to DIN EN ISO 5167-1
minimum inlet run 15 × pipe diameter D, outlet run 5 × pipe diameter D (see page
11
)
Pressure loss
(logarithmic diagram)
< 1.0 kPa (10 mbar), typical value 0.1 kPa (1 mbar)
Electrical power values
Universal power supply unit:
110...230 V AC/DC ± 10%(f = 48...62 Hz)
Low-voltage power supply unit:
24 V AC/DC ± 20%(f = 48...62 Hz)
Power dissipation
20 VA
15 VA
Connections
M20 x 1.5 or ½“ NPT
Cable (remote version)
Round signal line with braided copper shield LIYCY 10
x 0.5 mm
2
.
max. 25 m between transducer and electronics unit
Output signals
PROFIBUS DPV1 version
EN 50170, acc. to PA profile 3.0
Analog/HART version
Analog output
Digital output
Digital input
0/4...20 mA, load < 600
Ω
2 x passive optocoupler (approx. 100 mA) selectable as frequency, pulse, or switching output
2 x 24 V lin typ. 10 mA (low < 2 mA, high > 10 mA) frequency and switching input
10
50
100
500
1000
5000 10000
10
5
0,5
0,1
1
DN 50
DN 80
DN 100
DN 150
DN 25
Pressure drop (mbar)