Operation, configuration and communication
42/14-39 EN
Sensyflow iG, Version 1.00
33
7.1.3.5 Totalizer
Function Block Totalizer (TOT)
Slot
Index
Relative
index
Parameter
Object data
type
Store
Size
Access Transport
type
Default value/
Allowed range
Mandatory (M) /
Optional (O)
3
16
0
Block Object
DS-32
C
20
r
C/a
-
M
3
17
1
ST_REV
uint16
N
2
r
C/a
0
M
3
18
2
TAG_DESC
OctetString
S
32
r,w
C/a
'
M
3
19
3
Strategy
uint16
S
2
r,w
C/a
0
M
3
20
4
Alert_Key
uint8
S
1
r,w
C/a
0
M
3
21
5
Target_Mode
uint8
S
1
r,w
C/a
-
M
3
22
6
Mode_BLK actual
permitted
normal
DS-37
D
3
r
C/a
block-specific
M
3
23
7
Alarm_SUM
DS-42
D
8
r
C/a
0,0,0,0
M
3
24
8
Batch
DS-67
S
10
r,w
C/a
0,0,0,0
M
3
25
3
26
10
Total
DS-33
N
5
r
O/cyc
0
M
3
27
11
UNIT_TOT
uint16
S
2
r,w
C/a
Direct Integral
of the Channel
M
3
28
12
Channel
S
2
r,w
C/a
-
M
3
29
13
Set_Tot
Uint8:
0:Totalize
1:Reset
2:PRESET
N
1
r,w
I/cyc
0:Totalize
M
3
30
14
Mode_Tot
uint8 0:Balan-
ced
1:Pos_only
2:Neg_only
3:Hold
N
1
r,w
I/cyc
0:Balanced
M
3
31
15
Fail_Tot
uint8 0:Run
1:Hold 2:Me-
mory
S
1
r,w
C/a
0:Run
M
3
32
16
Preset_Tot
Float
S
4
r,w
C/a
0
M
3
33
17
Alaram_Hys
Float
S
4
r,w
C/a
0
M
3
34
18
HI_HI_LIM
Float
S
4
r,w
C/a
Max. Value
M
3
35
19
HI_LIM
Float
S
4
r,w
C/a
Max. Value
M
3
36
20
3
37
21
3
38
22
HI_HI_ALM
DS-39
D
16
r C/a
0
O
3
39
23
HI_HI_ALM
DS-39
D
16
r C/a
0
O
3
40
24
HI_HI_ALM
DS-39
D
16
r C/a
0
O
3
41
25
HI_HI_ALM
DS-39
D
16
r C/a
0
O
3
26-35
Reserved by PNO
M
3
52
36
Manufacturer
Integrator Time
String
S
4
R
A
-
Years,
Days,
Hours,
Minutes,
O