2101510MNA H | NGC820 0 AN D PGC 1000 US ER MA NUA L |
77
Figure 7-3: Calibration running
6.
When the calibration is complete, the device should move to the designated Next Mode. View
results on the Analyzer Operation screen.
7.
The un-normalized total of the calibration stream should be between 99.5% and 100.5%. If
values exceed these parameters, proceed to the
section.
8.
Carefully examine the calibration stream’s Chromatogram-1 and Chromatogram-2 by clicking
the
button. Look for unlabeled peaks and base line anomalies. Use
and
as a guide. If errors exist, proceed to the
section.
9.
Once the device is running smoothly, is producing good chromatograms, and all the peaks are
labeled and eluting correctly, perform a save and restore procedure to update tfCold.
Figure 7-4: Typical chromatograph for chromatogram-1 (heavies)