54
| NG C820 0 AN D PGC1000 US ER MAN UA L | 2101510MNA H
Figure 6-15: IP configuration for host system (automatic addressing)
6.3.3.3
Set up PCCU32 and connect
This procedure assumes that the default Ethernet configuration is intact. The analyzer has a default IP
address assigned at the factory (169.254.0.11).
To configure PCCU32 for TCP/IP communication:
1.
Connect an Ethernet cross-over cable between the PC and analyzer. Make sure the PC or laptop
is powered on.
2.
Power on the analyzer. Diagnostic tests begin immediately. See section
for details.
3.
Launch PCCU32.
4.
Click
Setup
on the PCCU32 toolbar menu.
Figure 6-16: PCCU32 Setup
5.
On the System
Setup
tab, under PCCU Connect Method, click
TCP/IP
6.
Type the default IP address (
169.254.0.11
) into
Network ID or IP
field.