IMPORTANT NOTE:
Note that the termination panel to digital controller ribbon cable pin 1 wire is
not red. On the digital controller board, the red edge (pin 1) of the cable should plug onto pin 50,
the right side of the plug. The plug is keyed; do not force plug into connector.
22. Insert the lithium battery pack into the enclosure between the enclosure and the thermal flask.
23. Once the device is reassembled, apply power to the analyzer.
24. Reinstall the front and rear end caps.
IMPORTANT NOTE:
Since the power was removed from this device, the analyzer will perform
startup diagnostics and stabilize. If the user has disabled the startup diagnostics, they should be
enabled and power cycled to the device. If the power has been withheld from the device for an
unknown or lengthy period of time, a complete startup should be performed.
For more information on enabling the diagnostics in PCCU, click
Diagnostics
and then
Help
.
10.25 Replace the analytical processor to the termination
panel cable
If the cable connecting the analytical processor to the termination panel requires replacement, use the
following instructions.
To replace the cable:
1.
On the Analyzer Operation screen, click
Hold
under Next Mode. When the device completes the
current cycle and enters hold, continue to the next step.
2.
Collect data from the device.
3.
Back up the configuration files following the instructions detailed previously in section
.
4.
Using the
instructions, verify the battery status is ok before proceeding.
5.
Disconnect or remove the power from the device externally, or remove the J1 connector from
the termination panel.
NOTICE – Equipment damage.
As with all electronic components, caution should be used
when handling the boards. Static electricity can potentially damage board components,
voiding any warranty.
6.
Following the instructions detailed in section
, remove the panel and unplug the cable.
Reaching into the enclosure, unplug the analytical processor to the termination panel cable
from the analytical processor assembly (see
).
7.
Using the replacement cable, insert it into the enclosure, and plug into the
power/communication connector, J1. Connect the cable to the back of the termination panel
J12 connector (see
).
8.
Reinstall the termination panel.
IMPORTANT NOTE:
Note that the termination panel to digital controller ribbon cable pin 1 wire is
not red. On the digital controller board, the red edge (pin 1) of the cable should plug onto pin 50,
the right side of the plug. The plug is keyed; do not force plug into connector.
9.
Once the device is reassembled, apply power to the analyzer.
10. Reinstall the rear end cap.
IMPORTANT NOTE:
Since power was removed from this device, the analyzer will perform startup
diagnostics and stabilize. If the user has disabled the startup diagnostics, they should be enabled
and power cycled to the device. If the power has been withheld from the device for an unknown or
lengthy period of time, a complete startup should be performed.
For more information on enabling the diagnostics in PCCU, click
Diagnostics
and then
Help
.