2101510MNA H | NGC820 0 AN D PGC 1000 US ER MA NUA L |
81
Figure 8-2: Straight through Ethernet cable pinouts
8.4
Enable Ethernet and configure IP parameters
This procedure changes the default configuration of the Ethernet interface. If you are performing this
procedure through a local connection on the Ethernet port, this connection will be lost after
configuration change. To test network connection after configuration change, make sure to connect
the device and laptop to the Ethernet switch or network router. The laptop must be reconfigured with
an IP address compatible with the new IP address assigned to the analyzer to reestablish connection.
IMPORTANT NOTE:
Analyzer reboot is required to enable the Ethernet interface. If the analyzer is
already in RUN mode, it will resume RUN mode after the reboot if the
Process Mode After Start-
Up
parameter on the Diagnostics tab is set to:
Previous Mode
.
You can also perform Ethernet configuration changes through a connection using either the USB or
MMI ports. If using the USB, establish a connection per section
, before you
start this procedure. It is assumed that the startup wizard was completed at first-time connection.
To enable Ethernet:
1.
Launch PCCU and click
Entry
. The analyzer’s Operation screen displays.
2.
Click
Show Tree View
in the upper left corner of screen (
).
Figure 8-3: Analyzer Operation screen
3.
On the navigation tree, click
Communications
. The Communication Ports screen displays
(
).