2101510MNA H | NGC820 0 AN D PGC 1000 US ER MA NUA L |
53
6.3.3.1
Cabling for connection to operator laptop
The analyzer’s Ethernet port does not have the ability to autoconfigure for cable type. Local operator
connection to the analyzer’s Ethernet port requires a crossover Ethernet cable.
shows
details.
Table 6-3: Ethernet cabling for local connection
Host system (PC/Laptop)
port type
Required cabling termination (connectors)
or adaptors
Maximum distance
Ethernet
Cross-over Ethernet CAT 5 cable with RJ-45
connectors at both ends.
100 meters (328 feet)
You can purchase crossover cables or build your own.
shows the crossover cable pinouts.
If you build your own cable, ensure the connectors are properly crimped and that no internal wires are
exposed outside of the connector. Test the cable for proper connectivity. If you are experiencing
connection issues, always check that you have the correct cable pinout and that the cable and
connectors are free from damage and are properly inserted in the ports.
Figure 6-14: Crossover Ethernet cable
6.3.3.2
Configure the host system
If the laptop is configured to automatically obtain the IP address, do not change the TCP/IP
configuration. If the laptop has a static TCP/IP configuration, configure it for dynamic IP addressing
and (automatic) private addressing (See
for Windows
®
7 system examples, screens may
differ depending on the Windows OS).