Figure 6-28: Save initial configuration
2.
Click
Send
and then
Next
to move to the next screen.
6.5.6
Analyze the calibration stream
This procedure determines if the analyzer provides accurate composition analysis of the calibration
gas blend. If composition analysis does not reflect the expected composition in the blend, the analyzer
requires calibration. Perform this procedure prior to running the process streams.
To run the analyzer’s calibration stream:
1.
At the
Analyzer Cal Stream
tab, select
Stream 4
on the left side of the screen (
The on-screen indicator next to
Stream 4
should light up, the cycle clock begins, and you
should see on-screen animation of gas flowing on the calibration stream (
).