56
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Test gases
Zero-point calibration
Nitrogen or air or UV sample component-free gas
End-point calibration
Calibration cells or test gas for each sample component
Dew point
The dew point of the test gases must be approximately the same as the dew
point of the sample gas.
NOTE
The notes for calibrating (see page 249) must be considered.
Pressure sensor
The pressure sensor is installed in the gas analyzer ex works.
The pressure sensor is located in the sample gas feed path if the internal gas
lines are executed as FPM flexible tubes. The connection of the pressure
sensor is led to the outside via an FPM hose if the internal gas lines are exe-
cuted as tubes. The connection of the pressure sensor is documented in the
pneumatic diagram contained in the analyzer data sheet.
Gas connections
see section "Limas11 IR, Limas21 UV, Limas21 HW: Gas connections" (see
page 84)