192
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Saving the configuration
Menu path
MENU
→
Configure
→
System
→
Save configuration
Automatic saving of the configuration
The database containing the configuration data and logbook recordings is
automatically saved in two configuration files.
The database is saved at the time when parameters have been changed in
the menu mode. Saving takes place either when the user deactivates an
entered password by pressing the "MEAS" key twice or when the gas analyz-
er automatically reverts to measurement mode by "time out".
When the gas analyzer is booting the last saved valid configuration file is
loaded.
Manual saving of the configuration
The database can also be saved manually. This is useful e.g. for buffering a
large function block configuration.
Backup
In addition to automatic or manual saving of the configuration a backup of
the current configuration can be created. This backup can be saved in a sep-
arate memory area and can be loaded when required e.g. to reset the gas
analyzer to a well-defined state.
NOTE
Using the software tool "SMT light" it is possible to save a backup of the
current configuration on a separate storage medium. "SMT light" can be
found on the DVD-ROM "Software tools and technical documentation" which
is delivered together with the gas analyzer.