![ABB AO2000 Series Скачать руководство пользователя страница 172](http://html.mh-extra.com/html/abb/ao2000-series/ao2000-series_operating-instructions-manual_4321172.webp)
172
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Changing a component
The (maximum four) components, which are displayed in the component list
(see page 165), can be changed. Either the parameters of a component can
be changed or one component can be replaced by another.
Menu path
MENU
→
Configure
→
Component-specific
→
Active
Component
Select component and confirm with
ENTER
.
Change parameter of a component
NOTES
In the case of a system component (Type S) only the parameter "Unit" can be
changed; in the case of a user component (Type B) all the parameters can be
changed with the exception of the parameter "Type".
The parameters of the active component (Type SA or BA) and the compo-
nent used for calibration (Type SK or BK) cannot be changed.
1
Press softkey
COMP. CONFIG
. The "Configure Component" menu is
displayed with the component list.
2
In the component list select the component with parameters which are
to be changed.
3
Press softkey
CHANGE COMP
. Enter password (level 3), as applicable. The
parameters of the component are displayed.