![ABB AO2000 Series Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/abb/ao2000-series/ao2000-series_operating-instructions-manual_4321038.webp)
38
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Climatic conditions
Relative humidity max. 75 %, no condensation
Ambient temperature for storage and transport
with integrated oxygen sensor
−
25 to +65 °C
−
25 to +60 °C
Ambient temperature range during operation with
Analyzer module
installed in a system
housing without
electronics module
installed in a system
housing with
electronics module or with
power supply only
Caldos25
+5 to +45 °C
+5 to +45 °C
Caldos27
+5 to +50 °C
+5 to +45 °C
Fidas24
+5 to +45 °C
+5 to +45 °C
Fidas24 NMHC
+5 to +40 °C
+5 to +40 °C
Limas11 IR
+5 to +45 °C
+5 to +45 °C
1)
Limas21 UV
+5 to +45 °C
+5 to +45 °C
1)
Limas21 HW
+15 to +35 °C
+15 to +35 °C
Magnos206
+5 to +50 °C
+5 to +45 °C
Magnos28
+5 to +50 °C
+5 to +45 °C
Magnos27
+5 to +45 °C
2)
+5 to +45 °C
Uras26
+5 to +45 °C
+5 to +40 °C
ZO23
+5 to +45 °C
+5 to +45 °C
Oxygen sensor
in 19-inch housing
in wall-mount h.
+5 to +40 °C
+5 to 35 °C
+5 to +40 °C
+5 to +35 °C
1)
+5 to +40 °C when I/O modules are installed
2) +5 to +50 °C for sample chamber direct connection and installation in
housing without electronics module or Uras26
Vibrations/shocks
If the gas analyzer is installed in a cabinet, the maximum acceleration am-
plitude may not exceed 0.01 ms
−
2
in a frequency range of 0.1 to 200 Hz.
If the gas analyzer is not installed in a cabinet, the following data for the
individual analyzer modules apply.
Analyzer module
Vibrations/shocks
Caldos25
max. ±0.04 mm for 5 to 30 Hz
Caldos27
max. ±0.04 mm for 5 to 55 Hz; 0.5 g for 55 to 150 Hz
Fidas24
max. 0.5 g, max. 150 Hz
Fidas24 NMHC
max. 0.5 g, max. 150 Hz
Limas11 IR
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
Limas21 UV
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
Limas21 HW
max. ±0.04 mm / 0.5 g at 5 to 150 Hz
Magnos206
max. ±0.04 mm at 5 to 20 Hz
Magnos28
max. ±0.04 mm at 5 to 20 Hz
Magnos27
max. ±0.04 mm at 5 to 60 Hz
Uras26
max. ±0.04 mm at 5 to 55 Hz; 0.5 g at 55 to 150 Hz;
slight transient effect on measured value in the region
of the beam modulation frequency
ZO23
max. ±0.04 mm at 5 to 55 Hz, 0.5 g at 55 to 150 Hz
Remark: Compliance with the metrological data can only be assured if data on
vibration amplitude and frequency range at the installation site are available
and suitable means are employed to decouple the analyzer.