![ABB AO2000 Series Скачать руководство пользователя страница 120](http://html.mh-extra.com/html/abb/ao2000-series/ao2000-series_operating-instructions-manual_4321120.webp)
120
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Adding an analyzer module to the system bus
1
Open the internal system bus connection between the existing analyzer
module and the electronics module in the gas analyzer.
2
Place a T-joint on the system bus connection (designated “BUS") of each
module (electronics and analyzer modules).
3
Connect the T-joints with the system bus cables.
4
Place a terminating resistor on the open ends of each T-piece.
5
Set up the added analyzer module.
Several analyzer modules: Connection via the system bus
BUS
PS
PS
EM
AM
AM
BUS
AM
Analyzer
modules
EM
Electronics
module
PS
Power
supply
BUS
System bus (external)
Terminating resistors