AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
177
Changing the unit of a component
The physical unit used for the measured value display of a component, e.g.
ppm or mg/m
3
, can be changed directly in the "Configure Component"
menu.
One of the units defined at the factory can be selected for system compo-
nents. For user components, the choice of units depends on which of the
parameters required for the calculation have been entered during configura-
tion (see page 168) of the component.
Changing a unit of component is possible for the following analyzer mod-
ules:
Analyzer module
Unit
Caldos27 ppm,
vol.%
Fidas24
ppmC1, ppm, vol.%, mgC/m
3
, gC/m
3
, mg/m
3
, g/m
3
,
%LEL, mgC/l, gC/l, mg/l, g/l
Magnos206 ppm,
vol.%
Magnos28 ppm,
vol.%
Magnos27 ppm,
vol.%
Uras26
ppm, vol.%, mg/m
3
, g/m
3
ZO23 ppm,
vol.%
Prerequisites for this are:
•
AMC Software version
≥
3.3.2,
•
Measured value display set to ppm at the factory,
•
Unit switching activated at the factory.
Menu path
MENU
→
Configure
→
Component-specific
→
Active
Component
Select component and confirm with
ENTER
.