Response factors for additional sample components
If a measuring component is added with a response factor which was not
factory-set, we recommend determining this as follows:
factor
Response
omponent
C
Component
Propane
Propane
Component
ion
concentrat
gas
Test
display
value
Measured
ion
concentrat
gas
Test
display
value
Measured
×
×
=
The following information should therefore be noted in particular:
•
The zero point should not differ essentially from the calibrated zero
point, in particular when measuring with small concentrations. Other-
wise the display must be offset against the deviation from the calibra-
tion of the zero point with the measured value display.
•
The response factors of a gas in nitrogen and in synthetic air may differ
considerably from each other.
•
The measurements should always be performed with a sample compo-
nent and a measuring range, e.g. TOC in ppm C1. When calculating the
test gas concentration, the number of C-atoms must be considered.
•
Test gas cylinders with high accuracy should be (1 % and better) used.
Other relevant variables
The following variables must be stored in the software of the gas analyzer
for each measuring component: molar mass, number of C-atoms, response
factor and strip factor. These variables are stored for the standard sample
components; they must be entered when adding a user component.
Molar mass
M
C
= 12.011 g/mol
M
H
= 1.008 g/mol
Molar volumes
V
m
= 22.414 for 0 °C and 1013 hPa
V
m
= 24.05 for 20 °C and 1013 hPa