178
AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Changing the unit
1
Press
COMP. CONFIG.
softkey. The "Configure Component" menu is
displayed with the component list.
2
In the component list select the component with unit which is to be
changed.
3
Press
UNIT SELECTION
softkey. Enter password (level 3), as applicable.
A list of the units available is displayed.
4
Select the desired unit and confirm with
ENTER
.
5
A security prompt appears with the information that when switching
the unit, the range limits of the component are automatically adjusted.
6
Confirm the switching of the unit with
ENTER
or cancel with
Back
.
7
Quit the menu with
Back
or
Meas
.
8
Check the range limits (see page 156) and the number of decimal places
(see page 158) and adjust if necessary.