AO2000
CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
193
Configuring status signals
Menu path
MENU
→
Configure
→
System
→
Status signals
Function
The status signal configuration is set at the factory per customer’s order.
Generally this configuration does not have to be changed in the field.
Selection
The following signals are available:
•
Individual status signals, i.e. Error, Maintenance request and
Maintenance mode
•
Overall status signal.
NOTES
If the status signal configuration is changed from "overall status signal" to
"individual status signal", the digital outputs DO2 and DO3 of the standard
function block application "Status signals/externally controlled calibration"
(see page 112) which possibly are assigned to limit value signals will be
overwritten with single status signals.
For further information on status signals, see the "System status: Status
signals" (see page 319) section.