Limas11 IR, Limas21 UV, Limas21 HW: Thermal link replacement
When should the thermal link be replaced?
A defective thermal link will usually be indicated by an insufficient sample
cell temperature
(T-Re.K)
or lamp temperature
(T-Re.L)
error message.
In this case check the thermal link and replace as necessary.
Thermal link replacement
2
1
CAUTION!
The sample cell and the lamp are hot (approx. 55 °C or approx. 60 °C)! Al-
low the modules to cool down after switching off the power supply (ap-
prox. 30 minutes).
1
Switch off the gas analyzer power supply!
2
Open the wall housing door or the 19-inch housing cover.
3
Disconnect the thermal link from the sample cell
1
and/or the lamp
2
.
4
Release the spring clips and/or retainer and pull the thermal link from
the opening.
5
Check the continuity of the thermal link; if necessary, insert a new ther-
mal link (part number 0745836) in the opening and secure it with the
spring clips and/or retainer.
6
Connect the thermal link.
7
Close the system housing so that it is tight.
Light penetration during operation leads to erroneous measurement
values and overranging ("Intensity" status message).
8
Turn on the gas analyzer power supply.