CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
No. Text
S A W F Reaction/Remark
254 The boot program of the ana-
lyzer is defective! Notify Ser-
vice!
x x Notify Service.
255 The program of the analyzer is
defective! Notify Service!
x x Notify Service.
Analyzer modules
300 No new measured values from
the analog/digital converter.
x x Notify Service.
301 The measured value exceeds
the range of the analog/digital
converter.
x x Check the sample gas concentration. Notify Service.
302 Offset drift exceeds half the
permissible range.
x Check analyzer module and sample preparation.
Permissible range: 150 % of smallest installed measure-
ment range; 50% of physical measurement range for
Uras26. Notify service when drift exceeds these values
303 Offset drift exceeds permissi-
ble range.
x x
304 Amplification drift exceeds half
the permissible range.
x Manually calibrate the indicated detector at zero and
end point. Check analyzer module and sample prepara-
tion.
Permissible range: 50 % the the detector sensitivity.
Notify Service when drift exceeds this value.
305 Amplification drift exceeds the
permissible range.
x x
306 The offset drift between two
calibrations exceeds the per-
missible range.
x These messages are generated by automatic calibration.
Check calibration for plausibility. Fix possible cause of
implausibility. Manually calibrate the indicated detector
at zero (No. 306) and end point (No. 307).
Permissible range: 15 % of the smallest installed meas-
uring range; 6 % of the smallest installed measuring
range for measurements on plants subject to approval
and pursuant to the 27th and 30th BImSchV (Federal
Regulation on Immissions)
307 The amplification drift between
two calibrations exceeds the
permissible range.
x
308 A computational error occurred
during the calculation of the
measured value.
x x Notify Service.
309 The temperature regulator is
defective.
x See the status message from the applicable tempera-
ture detector
310 Temperature correction was
turned off for this component
because the temperature
measured value is invalid.
x See the status message from the applicable tempera-
ture detector
311 The pressure regulator is de-
fective.
x x See the status message from the applicable pressure
detector
312 The
pressure
correction
turned
off for this component because
of invalid measured pressure
value.
x See the status message from the applicable pressure
detector
313 Cross-sensitivity
correction
is
impossible for this component
because the correction value is
invalid.
x See the status message from the applicable correction
detector
314 Carrier gas correction is im-
possible for this component
because the correction value is
invalid.
x See the status message from the applicable correction
detector
Auxiliary detector