Inspection and maintenance
CAUTION
The tasks described in this chapter require special training and under
some circumstances involve working with the gas analyzer open and
powered up. Therefore, they should only be performed by qualified and
specially trained personnel.
Inspection
Normal operation of the gas analyzer
In normal operation the measured values from the installed analyzer mod-
ules are shown on the display screen and the green "Power" LED is lit.
Periodic checks
Module, assembly
Function
Interval
Sample gas
flowmeter
Caldos25
10 to 90 l/h,
max. 200 l/h
Regularly
Caldos27
10 to 90 l/h
Fidas24
80 to 100 l/h
Fidas24 NMHC 80 to 100 l/h
Limas11 IR
20 to 100 l/h
Limas21 UV
20 to 100 l/h
Limas21 HW
20 to 90 l/h
Magnos206
30 to 90 l/h
Magnos28
30 to 90 l/h
Magnos27
20 to 90 l/h
Uras26
20 to 100 l/h
ZO23
5 to 10 l/h ± 0.2 l/h
Pneumatic module
disposable filter
Staining Regularly
Gas analyzer gas
lines
Seal integrity
Regularly
Seals between doors
and housing
Contamination, foreign material
Before each