91
13-5 Data Memory for Tare Value
The data memory function can store 20 sets of tare values for weighing.
The tare value in memory can be recalled easily using the MODE key and used for weighing.
The tare value in memory can be recalled and changed.
Notes
The recalled tare value can be changed, using the digital input mode or the weighing input mode.
The digital input mode enters the tare value using the keys. The weighing input mode uses a
sample tare container to store the tare value.
The NET indicator illuminates during tare operation.
Storing the tare value
To store a new tare value: Recall the stored tare value to be changed ("
t01
" to "
t20
"). Then, change the
recalled tare value using the digital input mode or the weighing input mode, and store the new value.
Notes
The recalled tare value can be changed, using the digital input mode or the weighing input
mode. The digital input mode enters the tare value using the keys. The weighing input mode
uses a sample tare container to store the tare value.
When the RE-ZERO key is pressed with nothing placed on the weighing pan, zero is
displayed, The NET indicator does not illuminate.
"
t--
" appears when a tare operation is performed without using the tare value stored in
memory.
While the data memory function is in use, unit selection using the MODE key is not
available.
1. Press the MODE key to select a unit to be used for storage.
2. Set the "Data memory (
data
)" parameter to "
4
".
3. Press and hold the PRINT key (for 2 seconds) until the
balance enters the tare value confirmation mode. The tare
value last selected is displayed.
4. Select the tare number to be used, using the following keys.
RE-ZERO key ............ To increase the tare number by one.
MODE key ................... To decrease the tare number by one.
CAL key ..................... To cancel it and to return to the weighing mode.
Each time the RE-ZERO key or MODE key is pressed, the displayed value changes as
follows:
t03
t04
........
t20
t01
5. T change the selected tare value:
To use the weighing input mode, press the SAMPLE key to enter the weighing input mode. Go to
"Weighing input mode" on the next page.
To use the digital input mode, press the SAMPLE key, then press and hold the MODE key (for 2
seconds) to enter the digital input mode. Go to "Digital input mode" on the next page.
No.3 tare number
Tare value