82
13-2
Data Memory For Calibration And Calibration Test
Characteristic
□
Calibration data (when and how it is performed) and calibration test data can be
stored in memory.
□
All the data in memory is available to be output at one time to a printer or
personal computer.
□
Up to 50 data sets of the latest sensitivity adjustment or sensitivity
calibration can be stored.
* When the memory capacity has been reached to 50, "
fUl
" " "
illuminates in order in the upper left of the display as shown below.
Storing the calibration and calibration test data
1. Set the "Data memory (
data
)" parameter to "
data 2
". Refer to "11. Function Table".
2. With the settings above, each time calibration or calibration test is performed, the data is stored
automatically.
Upper left of the display