100
Outputting the statistical results
1. Each time the MODE key is pressed, the display
changes: the results as selected in “Statistical function
mode output items (
5taf
)”, and ,
.
When pressing the SAMPLE key, the
previous item is displayed.
Note
□
When the number of data is 1, the coefficient of
variation
and relative error is displayed as - - - - - - - .
□
When the average is 0, the coefficient of
variation and relative error is displayed as - - - - - - - .
□
Statistical items are indicated on the upper
left of the display using the following
symbols.
2. When pressing the PRINT key while
displaying the statistical result, the statistical
result is output.
Symbol
Statistical item
Sum
Maximum
Minimum
Range (Maximum – minimum)
Average
Standard deviation
Coefficient of variation
Relative error of maximum value
Relative error of minimum value
Deleting the latest data
When the wrong data is entered, it can be deleted and excluded from
statistical calculation. Only the latest data can be deleted.
1. In the weighing mode, press the MODE key several
times to display
1
.
2.
Press the PRINT key to display
.
3.
Press the RE-ZERO key to display
.
4.
Press the PRINT key to delete the latest data and
exclude it from statistical calculation.
The number of data decreases by 1 when the balance
returns to the weighing mode.
Press several times
Output example
Function table
parameter
0
1
2
3
N 10
SUM
1000.00g
MAX 105.00g
MIN 95.00g
R 10.00g
AVE 100.00g
SD 2.800g
CV 2.8 %
MAX% 5.0 %
MIN% 5.0 %
Output the statistical result
Delete
the latest data
Delete
the statistical data
Display
the statistical result