![background image](http://html.mh-extra.com/html/aandd/gf-10202m/gf-10202m_instruction-manual_446426043.webp)
43
9-7
Correcting The Internal Mass Value Of The GX-M series (Auto)
Calibrate referring to "9-3. Calibration Using An External weight".
This is method of correcting the internal mass weight value based on an external weight.
After calibration with the external mass, the balance automatically loads and unloads
the internal mass and corrects the internal mass value.
The available masses are as shown in the table below. The corrected mass
value is maintained in non-volatile memory even if the AC adapter is removed
Model
Available mass
Factory setting
Adjustable range
GX-8202M
2kg to 8kg (1kg
interval
)
5kg
-99.99g to
+99.99g
GX-8202MD
2kg to 8kg (1kg
interval
)
5kg
GX-10202M
2kg to 10kg (1kg
interval
)
10kg
GX-12001M
5kg,
10kg
10kg
-999.9g to
+999.9g
GX-22001M
5kg,
10kg,
20kg
20kg
GX-32001M
5kg,
10kg,
20kg, 30kg
20kg
GX-32001MD
5kg,
10kg,
20kg, 30kg
20kg
Setting procedure
1. The internal mass value cannot be corrected at factory settings.
Refer to "9-6.Correcting The Internal Mass Value Of The GX-M
series" and enable changing of the function setting and
correction the internal mass value.
2. In weighing mode, press and hold the SAMPLE key to
display
.
3. Press the SAMPLE key several times until appears.
4.
If does not display, perform "1".
5. Press the PRINT key to display
Auto
.
6. When preparation is completed, press the PRINT key.
7.
is displayed and the internal mass value is automatically corrected.
8. When adjustment of the internal mass value is completed, is
displayed and calibration is performed automatically with the adjusted
internal weight.
9. When calibration is completed,
Auto
is displayed. Press the
CAL key twice to return to weighing mode.
10. Place the external weight used for calibration on the balance to check
whether the balance was corrected. If it is not corrected properly,
return to "2".
During
calibration
with
external
mass
Place
Replace
(for 2 seconds)
Press
several
times
Press and
hold