25
About the operation at when power is turned on
The balance will decide the reference zero point when the power is turned on (AC adapter is
connected).
Depending on the load condition at that time, it will automatically judge whether to perform zero or tare
operation. The condition for determining which is used is "power on zero range", and when power on
zero range is exceeded, the tare subtraction operation is performed.
About re-zero operation
By pressing the RE-ZERO key, the display can be changed to zero.
Re-zero with the RE-ZERO key will automatically determine whether zero or tare operation is
performed.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
subtraction operation is performed.
About measurement range
For the balance, the range that can be weighed is determined by model.
The total amount (net tare quantity) up to the maximum display of each model is displayed,
and when the maximum display is exceeded, E is displayed to indicate that the weighing range is
exceeded.
When in excess in negative, -E is displayed.
Model
Power on zero
range
Zero range
-E display range
GX-8202M,
GF-8202M
Approx.
±
800g
Approx.-800g to +160g
Approx.-800g or less
GX-8202MD, GF-8202MD
Approx.
±
800g
Approx.-800g to +160g
Approx.-800g or less
GX-10202M, GF-10202M
Approx.
±
1kg
Approx.-1kg to +200g
Approx.-1kg or less
GX-12001M, GF-12001M
Approx.
±
1kg
Approx.-1kg to +200g
Approx.-1kg or less
GX-22001M, GF-22001M
Approx.
±
2kg
Approx.-2kg to +400g
Approx.-2kg or less
GX-32001M, GF-32001M
Approx.
±
3kg
Approx.-3kg to +600g
Approx.-3kg or less
GX-32001MD, GF-32001MD Approx.
±
3kg
Approx.-3kg to +600g
Approx.-3kg or less