10-2 Initializing The Balance
This function returns the following parameters to factory settings.
□
Calibration data
□
Function table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
□
The data that is stored in the balance using the data memory function
□
External calibration weight
□
Function switch settings
□
Density of a liquid and a water temperature at density mode
Note
Be sure to calibrate the balance after initialization.
Setting procedure
1. Press the ON:OFF key to turn off the display.
2. While pressing and holding the PRINT key and the SAMPLE
key, press the ON:OFF key to display
p5
.
3. Press the SAMPLE key to display
Clr
.
4. Press the PRINT key.
To cancel this operation, press the CAL key.
5. Press the RE-ZERO key to change .
6. With displaying
Clr,
press the PRINT key to initialize the
balance. The balance will automatically return to the weighing
mode.
end
0
00
g
Clr
p5
W
it h p res s i ng
a
n d h ol di ng
H
ol d