109
Key operation
In case of setting, operate with the following keys.
Key
Function
Weighing value
(
gross
)
Operation
Zero setting
(
ZERO
)
Within the zero range
*1
Update a zero point and clear a tare
value.
Out of the zero range
*1
Do nothing
TARE
Plus value
Do tare and update a tare value
Gross zero
*2
(
Gross zero mark blinking
)
Clear a tare value
Minus value
Do nothing
*
1 "Zero range" means the range where the load is within
±
2% of the weight from the reference zero.
For the zero range for each model, reter to "6-1 Basic Operation".
*
2 "Gross zero" means the range where the minimum scale of gross (total amount) is zero in "g".
(The state in which the gross zero mark is lit.)
Note
To turn off the display of balance's display, press the ON:OFF key (Long press) for about 2
seconds.
Display
Mark
Description
NET
This lights when the tare is not zero.
G
This lights when the tare is zero.
PT
When the preset tare is set by the PT command, this lights together with the NET mark.
This lights when the minimum scale of the gross is in the range of zero in "g".
o
NET
G
PT
NET mark
G mark
PT mark
Gross zero mark