120
(2)
Method of measuring density (specific gravity) of solid
(
function setting
)
Note
□
Re-set the density of the liquid with "(3) Entering the density of a liquid" as necessary, such as
when the temperature of the liquid changes during measurement or when changing the type of
liquid.
□
In the density display, the 3 digits after the decimal point are fixed. The
minimum display cannot be changed by pressing the SAMPLE key.
Density measurement displays the density fixed by measuring the weight
in air and measuring the weight in liquid.
The relationship between each state and display is as follows.
Setting procedure
1. Check the weight measurement mode in air (g lights, blink ).
Press the RE-ZERO key to display zero without
placing anything on the weighing pan.
2. Place the sample on the weighing pan in air and
wait for the display to stabilize.
If outputting the mass of
the sample, press the PRINT key.
Next, press the SAMPLE key to fix the weight in air,
and move to the weight measurement mode in liquid (g
lights, blink ).
3. Transfer the sample from the weighing pan in air to
the weighing pan in liquid and wait for the display to
stabilize. If outputting the mass of the sample, press the
PRINT key.
Next, press the SAMPLE key to fix the weight in liquid
and shift to the density input mode (g turned off, lights ).
4. Enter the density of the liquid
.
Refer to "(3) Entering the density of a liquid" and set the
density.
Next, press the PRINT key to enter the density
mode. (g turns off, lights ).
5. If outputting the density, press the PRINT key.
If measuring another sample, press the SAMPLE key
and start with the weighing mode in air. The density unit
is "
".
6. Re-set the density of the liquid with "(3) Entering the
density of a liquid" as necessary, such as when the
temperature of the liquid changes during measurement or
when changing the type of liquid.
7. Press the MODE key to enter another weighing mode.
From liquid density input
Display the density
g
turns off, lights
Set the density in liquid
g
turns offs, lights
Weight in liquid
g
lights, lights
Weight in air
g
lights, blinks
Weight measurement mode in air
g
lights, blinks
Remove the sample
Move the sample from
the pan in air to the pan
in liquid
Place the sample on
the weighing pan in air
Содержание GF-10202M
Страница 141: ...141 29 External Dimention Unit mm ...
Страница 145: ...145 MEMO ...
Страница 146: ...146 MEMO ...
Страница 147: ......