27
Processing mark
Counting Mode Using The ACAI Function
The ACAI is a function that improves the accuracy of the unit mass automatically by increasing the
number of samples as the counting process.
ACAI: Automatic Counting Accuracy Improvement
After registering unit mass of "5", proceed to the following "7".
7. If a few more samples are added, the processing indicator turns on. To
prevent an error, add three or more. The processing indicator does not
turn on if overloaded. Try to add the same number of samples as
displayed.
8. The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch
the balance or samples on the pan until the processing indicator turns off.
9. Counting accuracy is improved when the processing indicator turns off.
10. Each time the above operation is performed, a more accurate unit mass will be obtained. There
is no definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the
same number of samples as displayed.
11. Remove all the samples used in ACAI and proceed with the counting operation using the
improved unit mass.
Note
ACAI will not function on the unit mass entered using the keys, or digital input mode.
Storing the unit mass
By using the data memory function, 50 instances of storing a
sample unit mass can be stored.
1. Set the function setting item "Data memory function ( ) " to
"Stores unit mass in counting ( ) " .Refer to "11.Function
Table".
2. The displayed
"
p
** "
is the selected unit mass registration
number.
3. Press and hold the PRINT key (for 2 seconds) to switch to the
mode to change the unit mass registration number.
RE-ZERO key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Changes the registration )
MODE key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Changes the registration number (-)
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Decides on the displayed registration number.
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Cancel the displayed registration number.
4. Multiple unit masses can be stored by registering them with different unit mass registration
numbers.
Note
*
p
**:
The
unit
weight registration number is entered.
*
Unit weight can be read by ″
UN
:mm ″ command.
(
mm corresponds to P01 to P50 with 01 to 50.
)
*
The read unit mass can output by
″
?UW
″ command and can be changed by
″
UW
″
command.
Note
*
ACAI cannot be used for the read unit mass.
Press and hold
(for 2 seconds)
+1
-
1
Содержание GF-10202M
Страница 141: ...141 29 External Dimention Unit mm ...
Страница 145: ...145 MEMO ...
Страница 146: ...146 MEMO ...
Страница 147: ......