![background image](http://html.mh-extra.com/html/aandd/gf-10202m/gf-10202m_instruction-manual_446426063.webp)
63
Entering the values
10. With displayed, press the PRINT key.
11. Display .
12. Press the PRINT key.
13. The current setting value is displayed with all of the
digits blinking.
14. When the current setting is not to be changed, press
the PRINT or CAL key to go to "15".
When the current setting is to be changed,
press the RE-ZERO
key and store the
following keys.
SAMPLE key ꞏꞏꞏꞏ Select the digit to change the value.
RE-ZERO key ꞏꞏꞏ Change the value of the digit selected.
MODE key ꞏꞏꞏꞏꞏꞏꞏ Switch the polarity.
PRINT key ꞏꞏꞏꞏꞏꞏꞏ Store the new setting and go to "15".
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Cancel the new setting and go to "15".
15.
When
CP LMt
is displayed, pressing the PRINT key
will display the currently set value.
If changing the setting value, it can be registered the
tolerance value with the following keys.
For tolerance value, enter the value with the reference value
set to 100%.
SAMPLE key ꞏꞏꞏꞏꞏꞏꞏ
Move the blinking digit.
RE-ZERO (+)key ꞏꞏ
Change the value of the blinking digit.
MODE (-)key ꞏꞏꞏꞏꞏꞏꞏ
Change the value of the blinking digit.
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
Register and go to "16"
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Cancel and go to "16"
16.
Press the CAL key twice to return to the weighing display.
x2
The current setting is not
to be changed.
The current setting is
to be changed.
or
Go to "12"