108
16.Gross Net Tare Function
Zero setting and taring can be operated separately, and data output for Gross (total amount), Net (net
amount), Tare (tare quantity) becomes possible.
When the gross net tare function is selected, the key operation is changed as follows.
Key
Operation
ON:OFF key
Zero setting
(
Operate as the ZERO
key)
RE-ZERO key
Tare
(
Operate as the TARE key
)
In order to use the Gross Net Tare Function, it is necessary to change the "setting of the Function
table".
16-1 Preparation Of Gross Net Tare Function
To use this function, enter the Function table as follow, and set "Application Function
ap fnc
" to "
4
" in
"Application mode
apf
". To return the normal weighing mode (Factory settng), set " Application mode
apf
" to "
0
".
Please set as follows.
Setting procedure
1. Press and hold the SAMPLE key
(for 2 seconds)
until
1
ba5fnc
of the function table is displayed, then release
the key.
2. Press the SAMPLE key several times to display
ap fnc
.
3. Press the PRINT key to display
.
4. Press the RE-ZERO key several times to
display .
5. Press the PRINT key to store the setting.
6. Press the CAL key to return to the calculating display.
Press and hold
Press several times
end
Press several times
(for 2 seconds)