For safety reasons the equipment must be switched off if there is a mains power
failure.
6.3.1
Worktops
The SATOVAC vacuum pump has been designed to work in a horizontal position, fitted on a rigid top
and parallel to the ground (or directly on the floor). Suitable worktops include service furniture
(where the operator is standing) or modelling tables (where work is done sitting down). All worktops
must be checked for stability. Should the pump be installed on machine or frames check that
anchoring is secure and complies with safety standards.
Make sure the machine remains at least
10-15 cm from walls so as to ensure
proper ventilation.
6.4
CONNECTING UP TO THE POWER SUPPLY
If the pump is operated by a three-phase motor, check that the direction of rotation
coincides with the arrow on the motor fan cover.
The user must provide a power connection with relative socket near the machine installation point.
The user must also install an adequate electrical circuit breaker upstream from the socket as well as
efficacious overload/indirect contact safety devices. Connection is effected via the safety plug (16A in
compliance with European standards) at the end of the power lead: this must be inserted in the mains
socket. When connecting check:
• that mains voltage and frequency are as indicated on the identification plate (incorrect power
supply voltage may damage the machine);
• that the mains supply has a proper, efficient earth connection.