86
Operation
G-30-1426-en
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009
Preparations for use
Setting an ergonomic working height
Move the surgical microscope to a convenient position before each sur-
gical procedure and set the optimum ergonomic working height of the
stand via the lifting column (without patient!).
As long as you keep switch (2) pressed down, the lifting column (1) in the
stand base moves upwards or downwards, depending on the position of
the switch. When you release the switch, the lifting column stops immedi-
ately.
The switching technology causes a dead time of approx. 2 seconds at the
upper and lower end positions of the lifting column. After expiry of this
time, you can move in the opposite direction again by activating switch
(2).
Note:
The lifting column moves the microscope in the OP position prior to the
operation.
Do not constantly move the lifting column up and down, since a thermal
cut-out then automatically switches off the drive motor. If this occurs, the
lifting column can be moved again only when the motor has cooled down.
Warning!
– Do not activate the lifting column during surgery!
– Do not use the lifting column for focusing.
– When using the fundus-imaging system which is installed between the
surgical microscope and the patient, make sure that the patient is nei-
ther put at risk nor injured:
– by the motorized movement of the lifting column,
– by the motorized focusing system,
– by the movement of the suspension arm.
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...