Adjusting the balance setting of the suspension arm
•
Now perform the balance setting procedure with the complete micro-
scope equipment attached!
Note:
Before you precisely adjust the balance setting of the suspension arm, we
recommend performing a coarse balance setting of the arm. For this, the
suspension arm must be locked in its horizontal position.
•
To perform the coarse balance setting, move the suspension arm
slightly up and down. At the same time, turn the adjustment screw (2)
until you think that the spring force is sufficient to compensate for the
weight of the surgical microscope and accessories.
Note:
Clockwise turning increases the spring force, counterclockwise
turning reduces the spring force.
•
Hold the suspension arm and pull out the locking knob (1). This must
be possible without requiring a special effort. Otherwise, re-adjust the
spring force using the adjustment screw (2).
•
Press one of the release keys for the magnetic brakes on the micros-
cope during the balance setting procedure. Move the suspension arm
alternately up and down by approx. 20 cm. Using adjustment screw
(2), adjust the spring force in such a way that the effort required to
move the arm up or down is the same in both directions.
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...