128
Operation
G-30-1426-en
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009
OPMI
®
VISU 150 / 200: selecting the magnetic brakes to be released
You can define which magnetic brakes of the suspension system are to
be released when the keys on the left and/or right handgrips (1) and (2)
are pressed.
The top section (SELECT BRAKES:) of the display and key field in confi-
guration mode 2 offers you the following options:
Selection
At each press of the "
∇
" or "
Δ
" key, the program jumps from one possible
configuration to another.
Saving parameter settings
As soon as you have entered a parameter setting, it is saved under the
current user ID.
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
XY - Z The suspension system’s magnetic brakes for the XY and Z
movements can be released separately by pressing key (1) or
(2) in the left or right handgrip respectively.
•
To release the magnetic brakes for XY movement only:
press the key in the left handgrip (1).
•
To release the magnetic brakes for Z movement only:
press the key in the right handgrip (2).
•
To release all magnetic brakes, i.e. those of the suspension
system for XY and Z movement:
simultaneously press the keys in the left and right hand-
grips (1) and (2).
ALL
All magnetic brakes are always released, i.e. those of the sus-
pension system for XY and Z movement, no matter whether
you press the key in the left or right handgrip (1) or (2).
Path: The basic mode is automatically displayed after the instrument has
been switched on.
To access configuration mode 2 from the basic mode, go via con-
figuration mode 1.
For this, simultaneously press the "MODE" and "STORE" keys
twice in succession:
1. Jump from the basic mode to configuration mode 1
2. Jump from configuration mode 1 to configuration mode 2.
1
2
Configuration
mode 2
MODE
STORE MANUAL
SELECT
BRAKES:
ALL
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...