24
Safety
G-30-1426-en
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009
Xenon illumination system
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
•
Replace the xenon lamp in good time.
•
Reset the service hour counter to "0" after replacing the lamp.
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
•
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
•
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
1
Switching to the backup lamp
The lamp module contains two xenon lamps. The second bulb is used
as a backup bulb which must be swung into the illumination beam path
if the first bulb fails.
•
If the xenon bulb fails, open the lamp module as follows: Press
button (7). The lamp module is slightly ejected.
•
Pull out the lamp module as far as it will go.
•
Turn knob (1) through 180° until it snaps in place. This moves the
backup lamp into the illumination beam path.
•
Push the lamp module all the way back into the lamp housing.
•
Reset service hour counter (6) to "0". Use a pointed object and press
it into the recess of reset button (5).
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...