146
Maintenance / Further information
G-30-1426-en
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009
Autofocus option
Problem
Possible cause
Remedy
See
Autofocus option is inopera-
tive
Camera control unit (CCU)
switched off.
Error code display: E20
•
Switch on MediLive
CCU.
-
Monitor switched off.
•
Switch on the monitor.
-
Failure of the autofocus op-
tion
•
Contact service dept.
-
Poor focusing result or none
at all.
Focusing of the video
camera not matched to fo-
cusing of the surgical micro-
scope.
•
Focus the microscope
image, then the video
image.
-
Measuring field (ROI) incor-
rectly positioned.
•
Correct the ROI position
in configuration mode 4.
Suspension system moved
during focusing.
•
Do not move suspension
system during focusing.
-
No object in focusing range. •
Reposition surgical mi-
croscope.
-
No object with sufficient
contrast in focusing range.
•
Reposition surgical mi-
croscope or focus manu-
ally.
-
Poor focusing result or none
at all.
Incorrect entry for camera
orientation (left/right), menu
item "AF-CCD Position".
•
Activate the Video-AF
mode and change the
video camera orienta-
tion.
Incorrect entry for the video
focal length in menu item
"AF-V-FOC.".
•
Activate configuration
mode 4 and enter the
correct focal length.
Video camera not correctly
oriented.
•
Turn video camera so
that video image corre-
sponds to main
observer's viewing di-
rection.
-
Focused video image on
monitor, but unfocused
image in surgical micro-
scope.
Setting of diopter rings on
eyepieces has been
changed.
•
Set diopter rings to zero
or to the correct value.
-
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...