Operation
G-30-1426-en
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009
Adjusting the limit of downward movement
The vertical lift of the suspension arm must be limited in such a way that
the patient's safety is also ensured when the microscope is inadvertently
lowered.
•
Loosen the adjustment screw (1) by a few turns.
•
Press one of the release keys for the magnetic brakes on the surgical
microscope and lower the microscope to a position where it can be fo-
cused on the surgical field (depending on the focal length of the objec-
tive lens) and where the safety distance to the surgical field is still
sufficient.
•
Turn the adjustment screw (1) clockwise as far as it will go.
•
Move the surgical microscope again to the lower limit and check the
safety distance.
Summary of Contents for S88
Page 6: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 19: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 19...
Page 23: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 23 1 2...
Page 29: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 29 1...
Page 31: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Safety 31...
Page 32: ...32 Safety G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 135: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 Operation 135...
Page 142: ...142 Operation G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 180: ...180 Index G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009...
Page 181: ...G 30 1426 en S88 Floor Stand with Lifting Column Issue 9 0 Printed on 28 01 2009 181...