Introduction
Humphrey Matrix User Manual
2660021138935 Rev. F 2015-11
1-14
WARNING: SERVICE or REPAIR to be performed by QUALIFIED,
AUTHORIZED PERSONNEL ONLY. There are NO USER SERVICEABLE PARTS
INSIDE the Humphrey Matrix instrument. Disassembly of the instrument
presents a possible ELECTRICAL SHOCK hazard and will VOID the warranty.
If the unit fails, contact CZM for instructions.
WARNING: REPLACEMENT PARTS and ACCESSORIES – Use only approved
replacement parts and accessories.
CAUTION: The appliance coupler is the main disconnect device of the instrument.
Position the instrument in such a way to have easy access to disconnect the
appliance coupler in case of an emergency.
Electromagnetic Compatibility (EMC)
Note: Essential Performance: To provide accurate visual field measurements.
Note: The Matrix needs to be installed and put into service according to the EMC
information provided herein.
Note: Portable and mobile RF communications equipment can affect medical electrical
equipment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Matrix is intended for use in the electromagnetic environment specified below. The customer or user of the Matrix should assure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Matrix uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Matrix is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
CAUTION: The use of accessories, transducers and cables other than those specified
may result in increased emissions or decreased immunity of the equipment.
CAUTION: The Matrix should not be used adjacent to or stacked with other
equipment.
CAUTION: DO NOT USE the instrument near other equipment that produces strong
magnetic fields (such as MRI). The video monitor performance may be adversely
affected.
Summary of Contents for Humphrey Matrix 800
Page 1: ...User Manual Humphrey Matrix Model 800...
Page 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Page 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Page 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Page 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Page 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Page 157: ......