Environmental Specifications
Operating Conditions
Operating Temperature: +15° C to +35° C [+59° F to +95° F]
Operating Humidity: 30% to 75% non-condensing
Operating Altitude: 700 hPa to 1060 hPa
Storage and Shipping Conditions
Storage Temperature: -20° C to +60° C [-4° F to +140° F]
Storage Humidity: 0% to 90% non-condensing (limited to 90% @ 40° C, 50% @ 50° C)
Storage Altitude: 700 hPa to 1060 hPa
Visual Field Test Specifications
FDT Test
Visual Field
Locations
Probability
Level
Classifications
Fixation
Catch
Trials
False
Positive
Trials
False
Negative
Trials
Test
Strategy
N-30-5 (-1)
Screening
19
4
3
3
0
Supra-Threshold
24-2-5 (-1)
Screening
55
2
10
10
0
Supra-Threshold
N-30-F
Threshold
19
4
6
6
3
MOBS
24-2
Threshold
55
4
10
10
6
ZEST
30-2
Threshold
69
4
10
10
6
ZEST
10-2
Threshold
44
4
10
10
6
ZEST
Macula
Threshold
16
4
6
3
0
ZEST
FDT Test
Field of view
tested
(degrees)
Approximate
Target size
(degrees)
Spatial
frequency
(cycles /
degree)
Temporal
frequency
(Hz)
N-30-5 (-1) Screening
Central 30
10 x 10
0.25
25
24-2-5 (-1) Screening
Central 30
5 x 5
0.50
18
N-30-F Threshold
Central 30
10 x 10
0.25
25
24-2 Threshold
Central 30
5 x 5
0.50
18
30-2 Threshold
Central 30
5 x 5
0.50
18
10-2 Threshold
Central 10
2 x 2
0.50
12
Macula Threshold
Central 5
2 x 2
0.50
12
Summary of Contents for Humphrey Matrix 800
Page 1: ...User Manual Humphrey Matrix Model 800...
Page 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Page 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Page 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Page 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Page 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Page 157: ......