Humphrey Matrix User Manual
2660021138935 Rev. F 2015-11
Error Codes
D-1
(D) Error Codes
Error codes that require additional user information have been included in this table. Those
that are self-explanatory have not.
Code
Number
Message
(SBC V06.00.0X)
Possible
Cause
Recommended
Action
115
Not enough data for Overview found.
Select a patient with enough data.
A minimum of 2 visual fields of the
same test type are required for an
overview plot.
A minimum of two visual fields of the
same test type are required for an
overview plot
123
Please limit input to
CHARACTER_LIMIT characters.
Delete characters until the number of
characters used is less than or equal
to the number specified.
148
The default folder
DEFAULT_FOLDER_NAME contents
will be merged but the default folder
cannot be deleted. Continue?
Merging any folder with the Main
(default) folder causes this message.
Select Continue.
151
Invalid folder name entered. Please
enter a valid folder name.
Empty folder name field.
Enter folder name.
160
Invalid file name entered. Please
enter a valid file name.
Empty file name field when saving
Overview analysis plot as JPEG.
Enter valid file name.
167
WARNING! Are you sure you want to
permanently delete the selected test
from the database?
User confirmation needed.
Deleting tests from the database is
permanent. Unrecoverable if not
available from a prior backup.
168
WARNING! Selected test will be
permanently deleted from the
database. Continue?
User confirmation needed.
Deleting tests from the database is
permanent. Unrecoverable if not
available from a prior backup.
175
Patient not found in database. Add
New Patient?
User confirmation needed.
The automatic search feature did not
find a patient with the name entered
in the database – either new patient
or a data entry error.
176
Unable to run test during calibration.
Wait until calibration is completed.
All testing operations are disabled
during calibration.
Check Calibration status in Help (F6)
Calibration screen.
Calibration takes approximately 15
minutes to complete.
SW error if message doesn't go away
when calibration is completed.
(restart).
192
Sample test is not present on the
system. Function unavailable.
• The file
/root/GUI/printSample.fdt2
is missing
• The file
/root/GUI/printSample.fdt2
could not be loaded for various
file I/O reasons.
Try again. If problem persists, send
customer update (from web site?)
which includes necessary file. If after
upgrade, problem still is present may
need service but if this is the case,
the customer should be seeing other
issues with the system as well.
196
What would you like to do with these
tests?
If the user highlighted the tests and
selects the Edit test option on the F3
screen
The user decides to either move the
selected tests to a different folder OR
move the test association to another
patient.
Summary of Contents for Humphrey Matrix 800
Page 1: ...User Manual Humphrey Matrix Model 800...
Page 26: ...Introduction Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 1 20...
Page 84: ...Viewing and Printing Tests Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 6 18...
Page 108: ...Networking Configuration Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 8 18...
Page 150: ...OfficeMate PMS Instructions Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 B 2...
Page 156: ...Error Codes Humphrey Matrix User Manual 2660021138935 Rev F 2015 11 D 4...
Page 157: ......