Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
The new configuration is read by the device (as a check) and a backup is
created.
Fig. 58 Backup of factory configuration
15
Open and transmit protocol
Click
Open Protocol
to open the acceptance protocol.
Fig. 59 Opening the acceptance protocol
This protocol can be printed out or sent in digital form to the customer.
Send the protocol to the Service department of Carl Zeiss Spectroscopy GmbH.
⇒