Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
9
Referencing – Calibration of cavity (black reference standard)
Place the black cavity on the sensor.
Click
OK
to start the calibration.
Fig. 50 Calibrating cavity (black reference standard)
10
Photometric verification – Cavity measurement (black reference standard)
Place the black reference standard on the centering aid.
Click
OK
to start the measurement.
Fig. 51 Photometric verification
Click
to repeat the measurement.
Click
to proceed to the next step.
= Repetition of last
part of program