Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
11
Insert lamp cable into cable clip
Insert the lamp cable
2
in the left cable clip
3
so that the connecting plugs
are secured between both cable clips.
Close the left cable clip again.
12
Attach screw-on cable tie
Screw on a new cable tie
1
. Use a TX 10 Torx screwdriver for this.
Tighten the cable tie sufficiently.
Fig. 34
2
3
1
Attaching the lamp cable
13
Close the housing
Once the lamp has been replaced, the device can be closed again.
Refer to chapter “4.6 Closing the housing” on page 37.
14
Calibrate the compensation data
After each lamp replacement, the various profile attachments must be
recalibrated.
Refer to chapter “5 Performing sensor calibration” on page 39.