EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
5 Performing sensor calibration
•
The halogen lamp has been replaced and the device is closed.
•
The “TestCenter” software is installed on your test computer.
If this is not the case, you can download the software from your download
portal at
•
The white and gray reference standard certificates are already on the test
computer. If this is not the case, they must be copied to the test computer.
•
If not already done so, put the following equipment ready:
– Laboratory power supply unit (000000-2045-548)
– Ethernet cable (000000-2036-573)
– Reference holder “Sapphire Flange“ (000000-1462-668)
– Reference holder “Sample pan - 5.5 mm thick base”
(in TURNSTEP) (000000-2130-893)
– Reference holder “Sample pan - thin base”
(in TURNSTEP) (000000-2285-627)
– Centering aid for reference standards (000000-2051-306)
– White reference standard with certificate (000000-2393-701)
– Gray reference standard with certificate (000000-2394-676)
– Black reference standard (cavity) (000000-1270-164)
1
Connect the device to the power supply
For acceptance testing with “TestCenter”, always connect the device to a
power supply unit.
2
Connect the device to the test computer
Connect the device to the measuring computer via Ethernet cable.
Make sure that you have the same IP address range. You may need to change
the IP address range on the test computer.
3
Apply the reference holder to the device
Apply the required reference holder to the device.
Requirements
Procedure