EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
4.6 Closing the housing
•
All service activities have been properly complete. All disconnected plug
connection have been reconnected and cables are laid properly.
•
Put the following spare parts ready:
– Housing seal (part of the Maintenance Package)
•
Put the following tools and equipment ready:
– Centering aid for seal replacement (part of the Service Package),
consisting of 8 mounting pins
– TX 25 Torx screwdriver
1
Apply the new housing seal and insert mounting pins
Always apply a new housing seal
2
to the housing
3
.
Make sure the seal is correctly positioned. The side of the seal with the larger
corner radius (red circle) must coincide with the corresponding corner of the
housing.
Insert the 8 mounting pins
1
into the threads of the housing
3
.
The 8 pins can be distributed freely to any of the threads as they will be
removed in a later step.
Fig. 41
2
3
1
Applying the new housing seal and insert mounting pins
Requirements
Procedure